Antech’s UKAS Calibration Services
UKAS CALIBRATION SERVICES

Schedule

Printer friendly versionPrintable Version
Electrical:
Parameter Measurement Generation Remarks
DC Voltage 0 to 60 kV 0 to 40 kV  
DC Current 0 to 2000 Amps

0 to 100 Amps

2 to 2000 Amps

Source simulated above 100 Amps
DC Power   20 W to 1 MW Simulated
AC Voltage

0 to 1100 V

0 to 50 kV peak

0 to 1100 V

0 to 40 kV peak

10 Hz to 1 MHz.
AC Current 0 to 2000 Amps

0 to 100 Amps

2 to 2000 Amps

Source simulated above 100 Amps
Sourced. Simulated 40 to 60 Hz
AC Power   20 W to 1 MW Simulated
Frequency DC to 1.2 GHz DC to 1.2 GHz  
Time 0.1s to 1 hour 0.1s to 1 hour Hand held stopwatch
Phase angle 0.1° to 359.9° -180.0° to +180.0°  
DC resistance 0 to 1 TΩ 0 to 100 GΩ  
AC resistance 1 mΩ to 10 KΩ 1 mΩ to 1 KΩ  
Capacitance 1 pF to 100 mF 50 pF to 1 F 100 Hz to 100 kHz
Inductance 0.1 mH to 30 kH 0.1 mH to 1 H 100 Hz to 100 kHz
       
Video PAL & NTSC   Pattern Generator
Video NTSC & PAL   Black Burst Vector Scope
    Analogue Video Vector Scope
    Digital video Vector Scope
    Wideband Video Vector Scope
Accelerometry:
Parameter Measurement Generation Remarks
Frequency step 10 Hz to 5 kHz 10 Hz to 5 kHz Charge, Bridge & Triaxial
Amplitude 0.1 gn to 10 gn 0.1 gn to 10 gn Charge, Bridge & Triaxial
Temperature:
Parameter Measurement Generation Remarks
Thermodynamic -80°C to 1100°C -80°C to 1100°C  
Electrical Simulation -270°C to 1768°C -270°C to 1768°C  
Infrared -15°C to 500°C -15°C to 500°C  
Air Temperature
Parameter Measurement Generation Remarks
Temperature in air -20°c to 100°c   Uncertainty: +/- 0.5°c
Humidity
Parameter Measurement Generation Remarks
Relative Measurement, Not Source atm
10% RH to 95% RH
   
Pressure:
Parameter Measurement Generation Remarks
Hydraulic pressure 0 to 5000 bar 0 to 5000 bar Gauge & Absolute
Gas pressure -1 to 500 bar -1 to 500 bar Gauge
  0 to 500 bar 0 to 500 bar Absolute
  0 to -1 bar 0 to -1 bar 'Vacuum'
Differential pressure 0 to 10 bar 0 to 10 bar Up to high static of 400 bar
  1 x 10‾5mbar to atmosphere High Vacuum
Torque:
Parameter Measurement Generation Remarks
All engineering units 0 to 1000 Nm    
Force:
Parameter Measurement Generation Remarks
All engineering units 0 to 5000 kgf    
Liquid Flow (Mineral Oils):
Volume:
Parameter Measurement Generation Remarks
Volume 0.6 to 25 l/min   Uncertainty: +/- 0.15%
Volume 25 to 2500 l/min   Uncertainty: +/- 0.1%
Volume 2500 to 3000 l/min   Uncertainty: +/- 0.13%
Mass:
Parameter Measurement Generation Remarks
Conventional mass 0.5 to 20 kg/min   Uncertainty: +/- 0.10%
Conventional mass 20 to 2000 kg/min   Uncertainty: +/- 0.075%
Conventional mass 2000 to 2400 kg/min   Uncertainty: +/- 0.12%
Oils:
Parameter Measurement Generation Remarks
All engineering units 2 to 200 cSt    
Liquid Flow (Water):
Volume:
Parameter Measurement Generation Remarks
Volume 0.6 to 4 l/min   Uncertainty: +/- 0.27%
Volume 4 to 600 l/min   Uncertainty: +/- 0.2%
Mass:
Parameter Measurement Generation Remarks
Conventional mass 0.6 to 4 kg/min   Uncertainty: +/- 0.27%
Conventional mass 4 to 600 kg/min   Uncertainty: +/- 0.2%
Gas Flow (Air):
Volume:
Parameter Measurement Generation Remarks
Volume 0.2 to 2 l/min   Uncertainty: +/- 0.23%
Volume 2 to 2000 l/min   Uncertainty: +/- 0.16%
Air Velocity:
Parameter Measurement Generation Remarks
All engineering units 0.29 to 30 m/s   Uncertainty: +/- 1.5% +0.08 m/s

Traceable to national standards, UKAS calibrations are as per UKAS schedule.